Murata DC voltmeter input-wire protection
Heating Rate Controller · PV1 / PV2 · DMR20-10-DCM-R-C · Existing reference: electrical issue R1.2
Manufacturer requirement
Murata's DMR20-10-DCM technical notes, section 4, specify an external, user-supplied 0.25 A / 300 V time-delay (time-lag) fuse for the supply wiring connected to the meter's TB1 (+) and TB1 (−) terminals, in accordance with applicable requirements. The meter's internal self-resetting fuse does not replace this external supply-wire protection.
The two-wire meters measure their own channel's external DC input. Their operating range is approximately 6–75 V DC; the heater system design envelope is 0–60 V DC. Murata specifies 18–22 AWG copper input conductors with suitable insulation; check its datasheet for terminal preparation, torque and installation requirements. These meters do not measure heater current.
Murata DMR20 installation datasheet — section 4 (page 2) →
Equipment and open actions
| Meter | Existing function | External protection | Disposition |
|---|---|---|---|
| PV1 / CH1 | External channel 1 DC supply voltage, TB1 (+/−) | NOT FITTED | Engineering selection, installation and test required |
| PV2 / CH2 | External channel 2 DC supply voltage, TB1 (+/−) | NOT FITTED | Engineering selection, installation and test required |
Required design and inspection checks
- Document the actual input-wire routes and DC source reference/earthing arrangements separately for PV1 and PV2; photograph and identify meter TB1 and upstream connection points.
- Select and document the 0.25 A / 300 V time-delay fuse(s), fuseholder(s), required conductor protection, DC-rated voltage/interrupt capability, compatible wiring and physical mounting.
- With both mains and external DC sources isolated and proved dead, implement the approved modification and apply durable wire IDs/protection labels.
- Verify conductor continuity, correct meter polarity, branch protection, clearances, strain relief, connection security and absence of shorts. Check DC source voltage and current-limiting behaviour.
- Perform the approved functional and fault/protection checks; record results, reviewer and authorisation before permitting heater operation.
- Issue a new controlled drawing revision and update the 70-route schedule and terminal records after verifying the actual modified wiring. Do not insert hypothetical rows into R1.2.
Corrective-action acceptance record
Blank fields are intentional. This page is not an electronic sign-off system; printing it does not constitute acceptance.
| Physical wiring and input protection selected | ______________________________________ |
|---|---|
| Fuse / holder model(s), rating, DC interrupt rating and quantities | ______________________________________ |
| PV1 fitted, inspected and tested | ______________________________________ |
| PV2 fitted, inspected and tested | ______________________________________ |
| Drawing / schedule issue and inspection report | ______________________________________ |
| Competent reviewer, date and site approval | ______________________________________ |
Status: OPEN until all required evidence is recorded and accepted.